Treatment of Blockage in Exhaust Port of Laibao Pump of ZYD-100 Oil Filter

2019/8/1??????view:

  SOGEVAC is equipped with built-in exhaust filters, which can trap oil mist even at a large gas flow rate and ensure that there is no oil mist in the exhaust gas. If the exhaust filter is blocked, the pressure relief valve (6/28) opens and the filter is bypassed. As a result, the oil content in the exhaust gas increases, and the fuel consumption of the pump increases. This problem can be solved by installing new exhaust filters (see Section 3.4).

  1. Laibao Pump Exhaust Blockage Warning:

  This situation must be solved by replacing the exhaust filter. In order to prevent oil mist from being discharged into the atmosphere beyond the limits stipulated in the regulations, it is the responsibility of users to maintain the equipment.

  In some cases, the necessity of installing exhaust pipes may be considered. Volatile substances can pass through filters. According to the nature of the process gas, we propose to connect an exhaust pipe, which is necessary when the exhaust gas is dangerous.

  2. Laibao Pump Exhaust Blockage Warning:

  Take safety measures suitable for your application and process gases.

  The outlet of the pump is also connected by internal threads (6/2), which can be connected to the hose through an internal threaded joint.

  The cross section of the outlet pipeline should be at least equal to that of the pump outlet. If the outlet pipe diameter is too small, there will be overpressure in the pump.

  Before installing the exhaust pipe, remove the flange plate of the exhaust outlet and ensure that the exhaust filter is fastened in the normal position. They sometimes loosen during shipment and installation. Loose exhaust filters generate smoke during pump start-up and operation.

  When installing the exhaust pipe, tilt it downward to prevent condensate flow from returning to the pump. If it can't be tilted, we strongly recommend installing a condenser. If several pumps are connected to the same exhaust pipe, ensure that the exhaust pipe has a suitable diameter and that check valves are installed at the outlet of each pump.

  3. Oil filter Lyboa pump warning:

  When the exhaust pipe is blocked or blocked, the pump shall not be started. Before starting, make sure that the blind plate or similar cut-off device on the exhaust pipeline must be opened and the exhaust pipe must not be blocked.

  Because such throttling will reduce the pumping speed, increase the temperature so that motor overload or dangerous overvoltage in the pump. Excessive pressure in the pump will damage the seal, make the oil mirror glass burst, or damage the pump body. In addition to explosive hazards, excessive back pressure can cause dangerous process gases to leak out of the pump. If the tank is blown with inert gas, the flow rate of inert gas should be limited. For these suggestions, please contact LEYBOLD.

  The air tightness of exhaust pipe must be guaranteed when dangerous gas is pumped by Lyboa pump of oil filter.


国产免费AV片无码永久免费| 国产av剧情md精品麻豆| 内射人妻无码色AV麻豆| 中文字幕日韩欧美一区二区三区| 亚洲中文无码| 另类尿喷潮videofree| 超h公用妓女精便器系列小说| 门卫老董| 丰满少妇弄高潮了www| 精品无码国产污污污免费网站| 亚洲A∨无码男人的天堂| 抱着h不拔出来h| 国产成人精品亚洲777人妖| 国产精品美女www爽爽爽视频| 日本a级毛片| 先锋影音人妻啪啪VA资源网站| 婷婷伊人久久大香线蕉av| chinese老太交70years| 少妇性活BBBBBBBBB四川| 成人午夜剧场| 捆绑白丝粉色jk震动捧喷白浆| a级国产乱理伦片在线观看| 美女裸体a级毛片| 被猛男伦流澡到高潮H| 蜜桃麻豆WWW久久国产精品| 淑娟两腿间又痒了| 国产免费AV| 夫妻二人体验交换夫妻的后果| 灌满了求你们停下np| 欧一欧二欧三| 性videos欧美熟妇hdx| 人人妻人人澡人人爽人人精品浪潮| 99re久久精品国产| 国色天香精品一卡2卡3卡| 粗糙的舌头刮着小豆| 8X福利精品第一导航| 拍裸戏时被C了H| 国产熟女一区二区三区| 无码人妻丰满熟妇区五十路百度| 国产69精品久久久久777| 日本韩国男男作爱GAYWWW|